Ultrameter
Operation
Manual
Model 4P
10 - 02 (WEB) EG
MYRON L
COMPANY
Przeglądanie stron 0
1 2 3 4 5 6 ... 19 20

Podsumowanie treści

Strona 1 - Model 4P

Ultrameter™ Operation ManualModel 4P10 - 02 (WEB) EGMYRON LCOMPANY

Strona 2 - ULTRAMETER

5. Temperature CalibrationTemperature calibration is not necessary in the Ultrameter.VII. MEMORYThis feature allows up to 20 readings with their te

Strona 3 - * ™ GE Corp

X. CALIBRATION INTERVALSThere is no simple answer as to how often one should calibrate aninstrument. The Ultrameter is designed

Strona 4

XII. TROUBLESHOOTING CHARTSymptom Possible Cause Corrective ActionNo display, even though Battery weak or not connected. Check connect

Strona 5

D. pH Buffer SolutionspH buffers are available for your other Myron L Company instruments inpH values of 4, 7 and 10. The Myron L buffer solutions ar

Strona 6

D. A Chart of Comparative Error:In the range of 1000 µS, the error using KCl on a solution that should becompensated as NaCl or as 442, is shown in t

Strona 7

measurements, and should be reset for different dilutions ortemperatures.C. When does it make a lot of difference?First, the accuracy of temp

Strona 8

XVII. GLOSSARY Anions - Negatively charged ions. See Solution Characteristics, pg. 26.Algorithm - A procedure for solving a mathematical problem.See

Strona 9

Ultrameter™OperationManualAddendumModels 4P & 6PSoftware Versions 2.03, 2.10, 2.51 & Later** See page 5 to determine the version of software o

Strona 10

Cell Check For Instruments with Software Version 2.10 & Later.In these versions, a Cell Check feature has been added to furtherincreas

Strona 11 - Figure 10

II. USER MODE CALIBRATION LINKA new function has been added to the Ultrameter that makes calibrationof the unit while in "User" mode easier,

Strona 12

7-11-00Instrument IllustrationMSMRMMYYRROONN LL CCOOMMPPAANNYY4PCONDRES TDS These 3 Measurement keys will:• Turn instrument on• Measure parameter•

Strona 13

4. Press the key until "User" is the only solution icon being displayed.5. Press the key. 6. The User mode calibration "

Strona 14 - Temperature

FEATURES and SPECIFICATIONSA. Features• Superior resolution 4 digit LCD displays full 9999 µS/ppm.• Accuracy of ±1% of reading (not merely full s

Strona 15

TABLE OF CONTENTSInstrument Illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1FEATURES and SPECIFI

Strona 16

XV. CONDUCTIVITY CONVERSION to TOTAL DISSOLVED SOLIDS (TDS) . . . . . . . . . . . . . . .26A. How it’s Done . . . . . . . . . . . . . . . . . . . .

Strona 17 - Ultrameter™

C. Operation of the Keys (See Instrument Illustration on page 1)1. Measurement Keys in GeneralAny of the 3 measurement keys in the upper part of th

Strona 18

B. Measuring ResistivityResistivity is for low conductivity solutions. In a cell cup the value may driftfrom trace contaminants or absorption fr

Strona 19

2. Disabling Temperature Compensationa. As in Procedure to Select a Solution, pg. 10, select “USER” mode.b. With “USER” selected, press . If

Strona 20 - Figure 9

c. Leaving CalibrationYou know you are finished when the “CAL” icon goes out. Pressing anymeasurement key abandons changes not yet accepted an